Mature cataract using AMO Signature Pro

Show Description +

Eva Liang, MD, manages a mature grade 4 cataract and demonstrates her standard nucleus disassembly technique.

Posted: 6/21/2018

Mature cataract using AMO Signature Pro

Eva Liang, MD, manages a mature grade 4 cataract and demonstrates her standard nucleus disassembly technique.

Posted: 6/21/2018

Please log in to leave a comment.

Comments

on nicolae-sebastian

6 years ago

Thank you !!!

Eva Liang

6 years ago

Here are my settings on the Signature Pro:<br />&bull; Bottle Height 75cm <br />&bull; For Sculpt: Aspiration 30 cc/min<br />&bull; Vacuum 55 mmHg<br />&bull; Phaco Power 80%<br /><br /> Venturi for Segment Removal:<br />&bull; Vacuum 390 MmHg<br />&bull; Phaco Power 80%<br /><br /> If I switch to peristaltic for better holding, those settings are:<br />&bull; Aspiration: 42 cc/min<br />&bull; Vacuum: 480 mmHg<br />&bull; Power: 70%

on nicolae-sebastian

6 years ago

What are your settings for this case please ?