Eva Liang, MD
Show Description +
Eva Liang, MD, manages a mature grade 4 cataract and demonstrates her standard nucleus disassembly technique.
Posted: 6/21/2018
Eva Liang, MD
Eva Liang, MD, manages a mature grade 4 cataract and demonstrates her standard nucleus disassembly technique.
Posted: 6/21/2018
Please log in to leave a comment.
Comments
Just Now
on nicolae-sebastian
5 years ago
Thank you !!!
Eva Liang
5 years ago
Here are my settings on the Signature Pro:<br />• Bottle Height 75cm <br />• For Sculpt: Aspiration 30 cc/min<br />• Vacuum 55 mmHg<br />• Phaco Power 80%<br /><br /> Venturi for Segment Removal:<br />• Vacuum 390 MmHg<br />• Phaco Power 80%<br /><br /> If I switch to peristaltic for better holding, those settings are:<br />• Aspiration: 42 cc/min<br />• Vacuum: 480 mmHg<br />• Power: 70%
on nicolae-sebastian
5 years ago
What are your settings for this case please ?